Essentials of European Medical Device Regulatory Affairs - June 2020
Vigilance
What is Vigilance ? In the context of Medical Devices :- – Vigilance is… “system for the notification and evaluation of INCIDENTs and FIELD SAFETY CORRECTIVE ACTIONS (FSCA)
involving MEDICAL DEVICEs” (ref. MEDDEV 2.12-1, rev. 8)
MDR contains details previously only detailed in MEDDEV 2.12-1 rev 8
109
Vigilance What must be reported? MDR Article 87 Manufacturers …… shall report, to the relevant competent authorities……. the following: (a) any serious incident involving devices made available on the Union market, except expected side-effects: - clearly documented in the product information and - quantified in the technical documentation and - subject to trend reporting (b) any field safety corrective action in respect of devices made available on the Union market, including: - FSCAs outside the EU if relevant to the CE marked device
Electronic reporting - EUDAMED database
110
Vigilance
Incident (article 2 (64))
Any malfunction or deterioration in the characteristics or performance of a device, made available on the market, including use-error due to ergonomic features, as well as any inadequacy in the information supplied by the manufacturer and any undesirable side-effect
Use Error (MEDDEV 2 12-1 rev. 8 Vigilance)
Act or omission of an act that has a different result to that intended by the MANUFACTURER or expected by the OPERATOR
111
37