Vigilance Field safety corrective action: (article 2 (68)) FSCA Corrective action taken by a manufacturer for technical or medical reasons to prevent or reduce the risk of a serious incident in relation to a device made available on the market Examples: - device return / exchange / destruction - Device Inspection /examination by the user - Device Modification (includes upgrades for software faults) - Changes to labelling/IFU Field safety notice (article 2 (69)) FSN A communication sent by a manufacturer to users or customers in relation to a field safety corrective action
155
Vigilance Reporting Timelines
Clock starts when manufacturer becomes aware Type of event When to report Maximum Serious public heath threat Immediately 2 days Death Immediately causal relationship established or suspected 10 days
Unanticipated serious deterioration in health
Immediately causal relationship established or suspected
10 days
Other Serious incidents Immediately causal relationship established (or reasonably possible)
15 days
‐
FSCA
Before implementing (unless FSCA is urgent)