Essentials of European Medical Device Regulatory Affairs
MDR- Software rule 11
If such decisions have an impact that may directly or indirectly cause: – the death or an irreversible deterioration of the state of health ‐ a serious deterioration of the state of health or a surgical intervention
Software intended to provide information which is used to take decisions with diagnosis or therapeutic purposes
Class III
Class IIa
Except
Class IIb
Software intended to monitor physiological processes.
‐ If it is intended for monitoring of vital physiological parameters, where the nature of variations is such that is could result in immediate danger to the patient.
Class IIa
Except
Class IIb
All other software
Class I
67
IVD classification
Is a Notified Body Required? Yes: design dossier review (includes compliance to common technical standards) Audit of Quality Management System Batch released by the NB Yes: Audit of technical documentation & quality management system
Annex II List A
HIV Hepatitis ABO blood grouping
Annex II List B
Rubella PSA* self test for blood glucose
Self Test
Pregnancy self test
Yes: review of design & labelling for lay user suitability
General
Test for hormones cardiac markers hematology and clinical chemistry tests
NO: Manufacturer self declares
* only cancer test listed
68