Module 14: Design Development and Certification of Medical Devices
17/11/2023
STERILISATION
ISO 14937: 2009 - Introduction: “Adventitious microbiological contamination of a medical device prior to sterilization should be minimized.
….Even so, medical devices could, prior to sterilization, have microorganisms on them, albeit in low numbers. Such medical devices are non-sterile ”
Masterclass: Lecture 11
The Organisation for Professionals in Regulatory Affairs
7
STERILISATION
ISO 14937: 2009 ● there is always a finite probability that a microorganism might survive regardless of the extent of treatment applied. – probability of survival is determined by the number and resistance of microorganisms and by the environment in which the organisms exist during treatment. ● Sterility of any one medical device cannot be guaranteed – Assessment is based on probability of a viable microorganisms being present
Masterclass: Lecture 11
The Organisation for Professionals in Regulatory Affairs
8