Module 14: Design Development and Certification of Medical Devices
17/11/2023
Verification / Validation
The Organisation for Professionals in Regulatory Affairs
25
Verification / Validation
User Need “New drug ‘X’ needs to be delivered via autoinjector”
User Req. 1.3 “The drug should be delivered into intramuscular tissue”
Clinical Study
Activation into air
Functional Req. 1.1.1 “The extended needle length shall be 20mm” (User Req. 1.3)
Functional Req. 1.1.2 “No drug shall be delivered from the needle until the tip has reached a maximum depth of 15mm” (User Req. 1.3)
Validation
Verification
The Organisation for Professionals in Regulatory Affairs
26