Module 19: Regulation of In Vitro Diagnostic Medical Devices
13/04/2023
Case Study #3
Device Test for the detection of chlamydia in urine or swab samples Event 5 samples negative with the assay but positive with another assay Potential clinical outcome?
Reportable? An event occurred
The device was a contributory cause The event could have led to harm
29
Case Study #3
Device Test for the detection of chlamydia in urine or swab samples Event 5 samples negative with the assay but positive with another assay
Action to be taken by user: Continue to use test but retest specific negative results. Consider the need for retest of stored samples and patient recall. The assay will be redesigned.
FSCA or FCA?
30